126 Instructions for use Isolette 8000 plus SW 5.n
Troubleshooting
11.3 System prompt messages
The following messages notify the user of conditions that are not hazardous but
require attention or correction.
They do not generate any audible or LED alarms.
11.3.1 General messages
11.3.2 Calibration messages
System prompt message Cause Remedy
Alarm Reset A previously latched alarm condi-
tion has been cleared with the
Audio Paused/Reset key.
Information only, no action
required.
Keypad Locked -
Press ©
The user presses a key while the
Keypad Lock key is illuminated
and active.
Unlock the keypad by pressing
the Keypad Lock key.
Performing Power-Up Tests This message is displayed after
power is supplied from the main
on/off switch and the system is
performing the self-test.
Information only, no action
required.
Procedural Silence This message is displayed when
no alarm conditions are present
and the Audio Paused/Reset
key is pressed.
Information only, no action
required.
System prompt message Cause Remedy
100% Cal The oxygen system is performing
100% calibration.
Information only, no action
required.
21% Cal The oxygen system is performing
21% calibration.
Information only, no action
required.
Cal Fail The servo control oxygen system
failed to calibrate.
Repeat the calibration proce-
dure. If the calibration procedure
fails again, refer the device to
service personnel.
Cal Pass The oxygen calibration was suc-
cessful.
Information only, no action
required.
Cal Required This message is displayed when
oxygen control is selected after a
power-up or after the sensor
module connector is discon-
nected from the incubator.
Calibrate the oxygen sensor.
Calibration Failed The scale fails the 5-kg calibra-
tion.
Retry calibration. If it fails again,
remove the scale from service.
Slide Out Sensor This message is displayed when
the Cal button is pressed during
oxygen calibration, but the sen-
sor module is not pulled out into
the calibration position.
Slide the sensor module out to
the calibration position.