144 Instructions for use Isolette 8000 plus SW 5.n
Reprocessing
NOTICE
Risk of damage to the device
The hood contains 4 tabs (one at each corner) which keep the deck securely in
place. Installing the upper cover when the hood is down may result in damage to
the upper cover and/or jamming of the hood in place.
Make sure that the hood is raised before attempting to install the upper cover.
8 Install the upper cover.
9 Install the T-bars on the upper cover.
10 Install the bed, X-ray tray, and scale (if installed).
11 Visually and physically examine the mattress for any holes or cuts that enable
the entry of fluids onto the inner foam. If the mattress is damaged, replace it.
CAUTION
Risk of injury and inability to maintain hygienic safety
If the incubator parts/assemblies are not reinstalled after cleaning or maintenance,
the hygienic safety of the device may be compromised.
Ensure the cover is installed on the mattress before reuse.
12 Install the mattress.
13 If equipped with the optional iris ports on the side access panels, install the new
disposable or reusable iris port sleeves:
A Install the smaller diameter elastic band of a new sleeve over the inner ring of
the port housing.