150 Instructions for use Isolette 8000 plus SW 5.n
Service
13.5.1 Maintenance instructions
Clean and disinfect device or components before each maintenance step, also
before returning for repair.
Water reservoir assem-
bly
2)
Every 2 years Exchange Users
Humidifier
2)
Every 3 years Exchange Users
Manifold
2)
Every 2 years Exchange Users
Heater/impeller cover
with duct cover
2)
Every 2 years Exchange Users
Tilt bar O-rings Every 6 years Exchange Service personnel
Collection bottle,
condensation
management system
2)
Weekly
3)
Exchange Users
Hose/plug assembly,
condensation manage-
ment system
2)
Weekly
3)
Exchange Users
Drain plug with O-ring
2)
Yearly Exchange Users
Maintenance:
Device servicing and
maintenance
Yearly Check and
maintain
Service personnel
Calibration:
O
2 Sensors
2)
Weekly
4)
Calibrate Users
Standard scales
2)
Every 1 year and
when necessary
Calibrate and
check accu-
racy
Users
OIML/NAWI scales
2)
Every 1 year and
when necessary
5)
Check accu-
racy
6)
Users
Adjustment:
Standard scales As required Adjust Users
OIML/NAWI scales As required
Adjust
6)
Specialized service
personnel
1) Replace if the material becomes brittle or sticky or if strips of material have become detached.
2) If installed.
3) And also with each new patient.
4) Calibration may need to be done more frequently if barometric pressure changes occur and/or highest level
of system accuracy is required by the clinician for patient treatment.
5) Measuring accuracy depends on local gravity, as determined by altitude and latitude. The specified accuracy
is only applicable if the scales have been calibrated to the installation site.
6) Verification of the OIML/NAWI scale may also be required by, and performed by, local metro-logical author-
ities.
Component Interval Measure Target group