Instructions for use Isolette 8000 plus SW 5.n 117
Troubleshooting
Low Skin Mode Disabled The controller is configured
for air temperature operation
but the skin temperature
mode is selected.
If skin temperature mode is
desired, select skin tem-
perature mode in the system
configuration menu (see
System configuration menu
on page 113).
Medium Stuck Key There is a malfunction of the
keys on the front panel of
the controller.
Switch off the incubator and
remove it from service.
Low MEDIBUS Comm. Error Communication has failed
after a successful communi-
cation.
Check the cable and the
connection.
To acknowledge the alarm,
press the Audio
paused/Reset key.
Contact DrägerService.
Alarm
priority
Alarm Cause Remedy