Instructions for use Isolette 8000 plus SW 5.n 173
Technical Data
15.3.2 Electromagnetic Immunity
15.3.3 Recommended separation distances from wireless
communication devices
To ensure that the functional integrity of this device is maintained, there must be a
separation distance of at least 1.0 m (3.3 ft) between this device and wireless
communication devices.
Immunity against Test level and required
electromagnetic environment
Electrostatic discharge (ESD)
(IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances
(bursts) (IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/output lines:
±1 kV
Impulse voltages (surges)
(IEC 61000-4-5)
Voltage, external conductor – external
conductor: ±1 kV
Voltage, external conductor – protective
ground conductor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
50 Hz: 30 A/m
Voltage dips and short interruptions in
the supply voltage (IEC 61000-4-11)
Voltage dips of 30 % to 100 %, 8.3 ms
to 5 s, different phase angles
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V, ISM bands: 6 V
Electromagnetic fields in the vicinity of
wireless communication devices
Various frequencies from 385 MHz to
5785 MHz: 9 V/m to 28 V/m