76 Instructions for use Isolette 8000 plus SW 5.n
Getting started
1 Ensure that the bed is level and not in the Trendelenburg or Reverse
Trendelenburg position.
The bubble on the scale level indicator is within the circle (see Symbols on
page 34).
2 Select display 2, and press the Weight button.
The weight screen is displayed.
3 If necessary, remove any objects from the bed.
4 Press the ->0/T<- button twice (see Buttons on page 42).
The weight display reads zero and the background color changes.
5 Place a weight of known value, but less than 7 kg (15 lb), on the bed and verify
that the correct weight is displayed.
6 Remove the weight from the bed, and press the Home button.
7.2.9 Utility shelf (optional component) functional check
1 Ensure that the utility shelf and support arm are secure.
2 Ensure that the utility shelf rotates with the desired resistance.