Care List for Julian Anesthesia Workstation
Only applies for non-infectious patients.
When used with infectious patients, all parts in contact with breathing gas must additionally be sterilized after
cleaning and disinfection.
114
What How often
Processing intervals
1)
with filter on
Y-piece
with filter on inspiratory
and expiratory port
without filter
Julian workstation
Front daily, other surfaces weekly
Vapor 19.n / Vapor 2000
Daily
Power cable, gas supply hoses,
equipotential bonding lead
Monthly
Breathing bag and hose
Daily Daily Per patient
Bellows
Weekly Weekly Daily
Cover of breathing system with
APL valve
Weekly Weekly Daily
Valve discs
Weekly Weekly Daily
Bottom part of breathing system
Weekly Weekly Daily
Container of bellows
Weekly Weekly Weekly
Valve discs (use a cassette)
Weekly Weekly Daily
Expiratory port
Weekly Weekly Daily
Absorber and insert
Weekly Weekly Daily
Filter of the sample gas recirculation
system
Weekly Weekly Weekly
Flow sensor
Weekly Weekly Daily
AGS housing
Weekly
AGS flow tube (without filter)
Weekly
Container for AGS buffer volume
Weekly
AGS transfer hose
Weekly
Gas disposal assembly tubing with
connector
Monthly
Waste gas connector with tube
Weekly
Hose of sample gas scavenging
Weekly
1)
The processing intervals depend on the use and position of the filters. The table is
merely intended as a guideline in the absence of established hospital protocols for
cleaning and disinfection intervals
Components which
can be processed