Selecting the Basic Screens
Selecting the Standard Screen Page
The standard screen page is automatically displayed
whenever a ventilation mode is started (MAN/SPONT,
CMV or PCV). The standard screen page can always
be selected during operation:
1 Press Q key
or
2 Press
S
key several times.
Display (example):
The most important parameters are grouped together
on the right-hand side of the screen.
Three waveforms are displayed on the left-hand side
(to select other traces, see page 77).
Measured values shown in gray
Julian performs Calibration automatically. The
measured values are shown in gray until sufficient
data is available and until the sensors have reached
their operating temperature.
In this way, Julian indicates that data accuracy is still
compromised when compared to the specifications in
the "Technical Data" section. The sensors can be
calibrated manually for maximum accuracy, see
page 85.
When measuring CO2, anaesthetic agent and N2O,
their numerical values are shown in grey during the
warm-up phase whenever the workstation is switched
on.
Because the gas measuring module switches off
internally in standby mode, calibration is carried out
each time the unit is switched on. This is indicated by
»Cal« in the measured value window.
The display CAL appears instead of the numerical
value during automatic calibration of a sensor
(e.g. O2 sensor: every 24 hours).
The flow sensor is automatically calibrated as soon as
a CO2 signal is detected by the gas measuring
system.
Julian takes account of the respective gas mixture
when calculating volumetric values.
Monitoring
Selecting the Basic Screen Pages
Operating Instructions Julian SW 3.n
66
SpO 2
CMV
SpO2
et
CO
2
38
98 67
6.0
10
MV
Freq
600 10 10 0251.0050
alarm
limits
auto set
limits
alarm
info
list
config.
O%
2
L/min
PEEP Freq. V T
mL
T:TIP I
%
P max
Fresh Gas O + N O2 2
curves
cmH
2
O
bpm
cmH
2
O
1:2.0
O
2
45 41
2
NO
53 51
Fi
Fet
Iso.
0.8 0.6
Enf.
0.6 0.5
MAC (age: 36)
1.3
T
I
:T
E
CO
2
40
0
PAW
0
20
Julian
2
1