Cleaning /Replacing Cooling Air Filters
Clean filters monthly
1 Remove all three cooling air filters from their holders.
Clean in warm water to which a detergent has been
added; dry thoroughly.
Insert cooling air filters into their holders without
creasing.
Replace cooling air filters after not more than one
year. Dispose of filters as domestic waste.
Replacing Water Trap
When fill level reaches the mark
or
when the message
CO2 LINE BLK ! is displayed
(if the sample line is correctly installed and free of any
blockage).
Hold water trap by its ribbed surface and pull it out of
its holder.
Push a new water trap into the holder until it clicks
into place
Dispose of old water trap as domestic waste.
Replacing O2 Sensor
When the message
FIO2 INOP! is displayed
or
when the sensor can no longer be calibrated.
1 Unscrew sensor holder containing O2 sensor.
2 Remove the spent O2 sensor from the holder and
insert a new O2 sensor.
1 Screw holder back in its place.
For disposal of spent O2 sensors, see page 120.