C-Lock-ECG synchronization (optional)
Requirements for ECG
synchronization signal Pos. pulse with voltage >4.5 V,
>10 ms duration for driving 2 mA
Max. permissible signal delay with
reference to current QRS complex 40 ms
Socket for 2-pin jack-plug,
dia. 3.5 mm
Jack layout
Signal isolation from other
electronic components
Dielectric strength 4 kV
Operating Data
Operating voltage 90 – 265 V, 45 - 65 Hz
Power input 400 W typically, max. 2.3 kW
Uninterruptible power supply With fully charged battery: at least 30 minutes
(auxiliary outlets are not powered!)
Line power fuse Automatic circuit-breaker 16 A
Auxiliary outlets Each with two fuses 2 A DIN 41622 110/120 V AC
Medical gas supply
O2 270 to 550 kPa (39 to 80 psi)
N2O 270 to 550 kPa (39 to 80 psi)
AIR 270 to 550 kPa (39 to 80 psi)
Max. cylinder pressure (O2 and N2O) 19,000 kPa (2,750 psi)
Drive gas consumption
Operation MV ±1 L/min AIR or O2
Standby and MAN/SPONT 0 L/min
Dimensions of Julian W x H x D 68 cm x 133 cm x 68 cm
Dimensions of the storage tray W x D 43 cm x 29 cm
Dimensions of the breathing system W x H x D 31 cm x 35 cm x 21 cm
Weight Approx. 90 kg (200 lbs)
(ready for operation without
Vapor units and gas cylinders)
Weight of breathing system without soda lime 4.8 kg
Screen Electro-luminescent screen, 10.4" diagonal
Protection class
Device I, type B conforming to IEC 60601-1
SpO2 sensor Type BF
electrically isolated from protective earth
Electromagnetic compatibility EMC Tested to IEC 60601-1-2
Technical Data
Interfaces
Operating Data
Operating Instructions Julian SW 3.n
143
m
Ground
Signal