SpO2 Measurement
(Available Option)
Sensor Selection
Select sensor according to the following criteria:
– Patient weight
– Patient mobility
– Possible application site
– Patient perfusion
– Duration of use
The following table lists the specific sensor types
available and their characteristics. It is intended as an
aid for selection.
1)
in unopened, undamaged packaging
Sensor type OXISENSOR OXISENSOR OXISENSOR OXISENSOR OXISENSOR
I-20 D-20 DS-100 A D-25 R-15
Age group Infants Children Adults Adults Adults
Patient 1 to 20 kg 10 to 50 kg >40 kg >30 kg >50 kg
weight
Period of Short and long- Short and long- Short-term Short and long- Short and long-
use term monitoring term monitoring monitoring term monitoring term monitoring
Patient Limited activity Limited activity Inactive Limited activity Inactive
mobility patients only patients only
Preferred Toe Finger Finger Finger Nose
measuring
point
Sterility
1)
Sterile-packed Sterile-packed ____ Sterile-packed Sterile-packed
WARNING !
Use only original Nellcor SpO2 sensors.
Follow Instructions for Use of the sensors –
incorrect positioning or use can cause tissue
damage.
Monitoring
SpO2 Measurement
Operating Instructions Julian SW 3.n
86