Setting Measuring Parameters
Select the line »parameters« with dial knob,
press to confirm.
The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO2 measurement (optional) on/off:
Select the line »SpO2 measurement« with dial knob,
press to confirm.
Select »on« or »off« with dial knob,
press to confirm.
Select sample rate for gas measurement:
Select the line »sample rate« with dial knob,
press to confirm.
Select »60« or »200 mL/min« with dial knob,
press to confirm.
Recommendation: Select 200 mL/min.
Set CO2 units of measurement:
Select the line »CO2 unit« with dial knob, press to
confirm. The last value selected is highlighted by a
gray background.
Select the desired units of measurement by turning
dial knob, press to confirm.
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
94
Standby / Configuration
calibration
Configuration menu for Measurement Parameters.
defaults
alarm sound
scale
parameter
record
interface
alarm limits
curves
basic configurations
CMV settings
PCV settings
Fresh Gas Flow Control
0 1 2 3 4 6 7 8 9 5
1 2 3 5 6 7 8 9 4
pls. tone
adult child
SpO
2
measurement
Sample rate
CO
2
unit
CO
2
scale
Pressure unit
Flow display
MAC-config.
on off
60 200 mL/min
mmHg kPa Vol%
0-75 0-110 mmHg
Pa*100 mbar cmH
2
0
MV Vt

Table of Contents

Other manuals for Dräger Julian

Related product manuals