Select the desired position with the dial knob and
press to confirm.
Display (example):
The waveform is replaced and the cursor returns to
the zz
zz
symbol in the selection menu.
To return to the standard page:
Press the dial knob,
or
press the Q key.
If a waveform field shall remain blank:
Select »blank« with the dial knob and press to
confirm.
Using the Volumeter Function
To observe and assess ventilation in manual and
pressure controlled ventilation modes and during
spontaneous breathing.
Upper bar:
Current tidal volume (VT), preceded by its numerical
value.
The bar graph shows generation of VT
Lower bar:
Volumeter (minute volume measurement) with a mark
between each breath, preceded by its numerical
value.
The scale of the bar display depends on the default
setting, see "Configuring During Operation, Selecting
the scale" on page 78 and "Configuring in Standby
Mode, selecting the scale" on page 93.
Starting the volumeter
Press dial knob.
Values are deleted and volumeter is restarted if the
dial knob is pressed again before 60 seconds have
elapsed.
Time elapsed in seconds is displayed above the bar
with the integrated volume on the left. Individual
breaths are separated by segments in the bar. The
volumeter stops automatically with an acoustic signal
after 60 seconds.
Measured values are displayed for 4 minutes and then
disappear.
Setting
Tidal volume VT
in L
Minute
volume MV
L/min
Scale for infants
and small children
0.2 2
Scale
for adults
1.0 10
Monitoring
Using the Volumeter Function
Operating Instructions Julian SW 3.n
76
MAN/SPONT
98 67
4.0050
SpO2
alarm
limits
CO -al.
2
off/on
alarm
info
list
curves
config.
Fresh Gas O + N O
22
O %2
L/min
VT
0.35
0
0.5
1
et
CO
2
38
6.0
10
MV
Freq
G
_
Volumeter running. To restart,
confirm !
Volumeter 46 s
4.1
05
10
Pleth.
O
2
45 41
2
NO
53 51
Fi
Fet
Iso.
0.8 0.6
Enf.
0.6 0.5
MAC (age: 36)
1.3
CO
2
40
0
CMV
98 67
600 10 10 0251.0050
To select curve position, turn and confirm !
SpO2
alarm
limits
auto set
limits
alarm
info
list
config.
O%
2
L/min
PEEP
cmH2O
Freq.
bpm
V T
mL
T:TIP I
%
P max
cmH2O
Fresh Gas O + N O2 2
curves
et
CO
2
38
6.0
10
MV
Freq
Pleth.
T
I
:T
E
Flow
Volumeter
CO
2
1:2.0
O
2
45 41
2
NO
53 51
Fi
Fet
Iso.
0.8 0.6
Enf.
0.6 0.5
MAC (age: 36)
1.3
CO
2
PAW
Pressures
Pleth
O
2
AGas
Flow
FG/volu
blank
Volumeter
FG/press
VT
0.35
0
0.5
1
Volumeter 46 s
4.1
05
10

Table of Contents