Manual Calibration
Julian calibrates its sensors automatically during
operation. The message CAL appears instead of the
current measured value during a calibration process.
Calibration of the O2 sensor and zero calibration of
the other gas sensors can be repeated manually, for
instance if automatic calibration was unsuccessful.
● Turn dial knob to select the »calibration« column,
press to confirm.
● Select and confirm the appropriate line with dial knob,
for example:
»O2-Sensor
21 vol.%«
The calibration process is started as soon as the
selection is confirmed. The clock symbol UU
UU
appears
at the end of the respective line, indicating that
calibration is in progress.
When calibration is complete, a check mark (✓)
appears instead of the clock symbol. A question
mark (?) indicates that calibration must be repeated
because the previous calibration attempt was not
successful.
Other settings can already be made during
calibration.
Monitoring
Manual Calibration
Operating Instructions Julian SW 3.n
85
CMV
600 10
10
0251.0050
O%2
L/min
PEEP
cmH O
2
Freq.
bpm
V T
mL
T:TIP I
%
Pmax
cmH O
2
Fresh Gas O + N O22
PAW
40
20
0
O
2
-Sensor
21 Vol.%
gas sensor
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult child
alarms
1:2.0
settings
✓
✓
T
I
:T
E
default
CO
2
on off
SpO
2
on off
HLM on off
FiCO
2
5
MAC-config.
MAC display
on off
MAC correction
on off
age 36
O -sensor 2
21 Vol.%
gas
sensor
CMV
600 10 10 0251.0050
✓
O%2
L/min
PEEP
cmH O
2
Freq.
bpm
V T
mL
T:TIP I
%
Pmax
cmH O
2
Fresh Gas O + N O22
PAW
40
20
0
calibration
1:2.0
T
I
:T
E