Emergency Startup
NOTE: This procedure should only be used when
there is no time for the self test!
1 Switch Julian on.
If necessary:
2 Adjust the Safety O2 flow control to a sufficient flow
and ventilate the patient manually .
Check Vapor setting.
Wait for the software to be internally loaded and for
the electronics to be self tested.
After approximately 40 seconds, the checklist
appears on the screen.
● Confirm checklist:
Press dial knob.
After a few seconds, the »Cancel test« menu key is
displayed on the right-hand edge of the screen
Display:
● Press the »Cancel test« menu key.
The system now only performs a minimal test.
Julian is ready for operation about 1 minute after
switching on. Calibration of the O2 sensor is
complete after about 2 minutes.
NOTE: Leak test and compliance test are not
performed. Accuracy levels specified under
"Technical Data" can therefore not be guaranteed.
NOTE: To prevent abuse of this feature, the self test
can only be cancelled 10 times in succession.
The self test cannot be cancelled the next time that
Julian is started and a complete self test must be
performed.
The system switches to standby mode after
completing the minimal test.
Display (example):
To switch the standby screen on again:
● Press the dial knob or any other key.
To start Julian:
● Close the Safety O2 flow control
● Select a ventilation mode, see "Operation",
pages 47 to 55.
Preparation
Emergency Startup
Operating Instructions Julian SW 3.n
44
Self-test
electronics
freshgas
mixer
ventilator
✓
cancel
test
36
delete
trend
leak
test
config.
Julian
03.00
01. May 00
Standby
2.00
L/min
100
Operate via either:
, or
MAN/
SPONT
CMV PCV
Fresh Gas O2
O2 %
Self Test interrupted 1x
Julian
D
1
2