Emergency Startup
NOTE: This procedure should only be used when
there is no time for the self test!
1 Switch Julian on.
If necessary:
2 Adjust the Safety O2 flow control to a sufficient flow
and ventilate the patient manually .
Check Vapor setting.
Wait for the software to be internally loaded and for
the electronics to be self tested.
After approximately 40 seconds, the checklist
appears on the screen.
Confirm checklist:
Press dial knob.
After a few seconds, the »Cancel test« menu key is
displayed on the right-hand edge of the screen
Display:
Press the »Cancel test« menu key.
The system now only performs a minimal test.
Julian is ready for operation about 1 minute after
switching on. Calibration of the O2 sensor is
complete after about 2 minutes.
NOTE: Leak test and compliance test are not
performed. Accuracy levels specified under
"Technical Data" can therefore not be guaranteed.
NOTE: To prevent abuse of this feature, the self test
can only be cancelled 10 times in succession.
The self test cannot be cancelled the next time that
Julian is started and a complete self test must be
performed.
The system switches to standby mode after
completing the minimal test.
Display (example):
To switch the standby screen on again:
Press the dial knob or any other key.
To start Julian:
Close the Safety O2 flow control
Select a ventilation mode, see "Operation",
pages 47 to 55.
Preparation
Emergency Startup
Operating Instructions Julian SW 3.n
44
Self-test
electronics
freshgas
mixer
ventilator
cancel
test
36
delete
trend
leak
test
config.
Julian
03.00
01. May 00
Standby
2.00
L/min
100
Operate via either:
, or
MAN/
SPONT
CMV PCV
Fresh Gas O2
O2 %
Self Test interrupted 1x
Julian
D
1
2

Table of Contents

Other manuals for Dräger Julian

Related product manuals