Setting Monitoring Functions
● Select »settings« column by turning dial knob,
press to confirm.
Display (example):
Setting volume of the pulse tone
0 = off
9 = maximum volume
● Select the line »pulse tone« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 5).
● Adjust value with dial knob, press to confirm.
Setting audible alarm volume
1 = minimum volume
9 = maximum volume
● Select the line »alarm sound« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 4).
● Adjust value with dial knob, press to confirm.
NOTE: Local standards require a volume of at least
45 dB(A) for audible alarms. Therefore, settings
1 through 4 are configured to deliver this audio level.
Selecting the scale
For scaling volumeter function and trend trace.
● Select the line »scale« with dial knob,
press to confirm.
● Select »adult« or »child« with dial knob,
press to confirm.
WARNING !
Always set audible alarm volume to a level
that is sufficiently loud with respect to the
general noise level of the surroundings.
Monitoring
Configuring During Operation
Operating Instructions Julian SW 3.n
78
CMV
600 10 10 0251.0050
O%2
L/min
PEEP
cmH O
2
Freq.
bpm
V T
mL
T:TIP I
%
Pmax
cmH O
2
Fresh Gas O + N O22
PAW
40
20
0
O
2
-Sensor
21 Vol.%
gas sensor
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult child
alarms
1:2.0
✓
✓
T
I
:T
E
default
CO
2
on off
SpO
2
on off
HLM on off
FiCO
2
5
MAC-config.
MAC display
on off
MAC correction
on off
age 36
settings