Setting Monitoring Functions
Select »settings« column by turning dial knob,
press to confirm.
Display (example):
Setting volume of the pulse tone
0 = off
9 = maximum volume
Select the line »pulse tone« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 5).
Adjust value with dial knob, press to confirm.
Setting audible alarm volume
1 = minimum volume
9 = maximum volume
Select the line »alarm sound« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 4).
Adjust value with dial knob, press to confirm.
NOTE: Local standards require a volume of at least
45 dB(A) for audible alarms. Therefore, settings
1 through 4 are configured to deliver this audio level.
Selecting the scale
For scaling volumeter function and trend trace.
Select the line »scale« with dial knob,
press to confirm.
Select »adult« or »child« with dial knob,
press to confirm.
WARNING !
Always set audible alarm volume to a level
that is sufficiently loud with respect to the
general noise level of the surroundings.
Monitoring
Configuring During Operation
Operating Instructions Julian SW 3.n
78
CMV
600 10 10 0251.0050
O%2
L/min
PEEP
cmH O
2
Freq.
bpm
V T
mL
T:TIP I
%
Pmax
cmH O
2
Fresh Gas O + N O22
PAW
40
20
0
O
2
-Sensor
21 Vol.%
gas sensor
calibration
pulse tone
alarm sound
scale
parameters
record
activate monitor defaults
0 1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
adult child
alarms
1:2.0
T
I
:T
E
default
CO
2
on off
SpO
2
on off
HLM on off
FiCO
2
5
MAC-config.
MAC display
on off
MAC correction
on off
age 36
settings