Setting the measuring parameters
Select the line »parameters« with dial knob,
press to confirm.
The »parameters« menu appears in the left-hand
field.
Display (example):
Switch SpO2 measurement (optional) on/off:
Select the line »SpO2 measurement« with dial knob,
press to confirm.
Select »on« or »off« with dial knob,
press to confirm.
Select sample rate for gas measurement:
Select the line »sample rate« with dial knob,
press to confirm.
Select »60« or »200 mL/min« with dial knob,
press to confirm.
Recommendation: select 200 mL/min.
Select CO2 units of measurement:
Select the line »CO2 unit« with dial knob,
press to confirm. The last value selected is
highlighted by a gray background.
Select desired units of measurement by turning
dial knob, press to confirm.
Setting pressure unit:
Select the line »Pressure unit« by turning the rotary
knob and press to confirm.
Select »Pa*100« or »mbar« or »cmH2O« by turning
the rotary knob and press to confirm.
Setting flow display:
Select the line »Flow display« by turning the rotary
knob and press to confirm.
Select »MV« or »Vt« by turning the rotary knob and
press to confirm. The selected parameter is displayed
on the standard page. Vt alarm limits are not
monitored. The alarm limits set for MV continue to be
monitored in operation.
Monitoring
Configuring During Operation
Operating Instructions Julian SW 3.n
79
CMV
0 1 2 3 4 6 7 8 9 5
1 2 3 5 6 7 8 9 4
adult child
600 10 10 0251.0050
O%2
L/min
PEEP
cmH O2
Freq.
bpm
V T
mL
T:TIP I
%
P max
cmH O2
Fresh Gas O + N O22
PAW
40
20
0
settings
pls. tone
alarm sound
scale
parameter
record
activate monitor defaults
Configuration Menu for Measurement Parameters.
anesth. agent calibration
alarms
1:2.0
T
I
:T
E
SpO
2
measurement
Sample rate
CO
2
unit
CO
2
scale
Pressure unit
Volume display
on off
60 200 mL/min
mmHg kPa %vol.
0-75 0-110 mmHg
Pa*100 mbar cmH
2
0
MV Vt