Installing the bellows
● Hold breathing system from above to prevent it from
dropping out and
1 uniformly slide the tightly compressed bellows onto
the connector.
2 Insert bellows container into the breathing system
from below and turn it clockwise as far as possible.
Filling and installing the absorber unit
3 Push insert fully into the absorber.
● Fill the absorber with fresh soda lime
– up to the MAX mark.
We recommend using Drägersorb.
If the breathing system is to remain unused during the
next 24 hours:
● Only fill with soda lime immediately before use!
Soda lime dries out when exposed to ambient air
and sun for a longer period of time. This reduces
the CO2 absorption capacity and the effectiveness
of the indicator.
4 Insert absorber into the breathing system from below
and turn clockwise as far as possible.
5 Slide in breathing system – as far as possible.
6 Fold down the writing top.
For installation of breathing circuits, suction system, etc.,
see page 27, "Preparation".
WARNING !
Absorbent is caustic and is a strong irritant to
the eyes, skin, and respiratory tract. When
filling absorber, take care not to spill its
contents.
Care
Assembling
Operating Instructions Julian SW 3.n
117
MAX
3
4
2
1
5
5
6