Adjustment range for default alarm limits:
Fixed alarm limits / times:
Apnea pressure after 15 seconds
Apnea flow after 15 seconds
Apnea CO2 after 15 seconds (after 60 seconds
in MAN/ SPONT)
– – : The alarm limit is deactivated as factory default.
NOTE: The newly set default alarm limits will be in
effect whenever the system is switched on or started
up from standby mode.
Certain alarm limits are switched off automatically in
MAN/SPONT: see tables, alarm limits for
MAN/SPONT, page 49.
Alarm limits Adjustment range
Factory default
SpO
2
[%]
_
51 to 100
50 to 99
– –
92
Pulse ª
[1/min]
_
31 to 300
30 to 299
120
50
EtCO2
[mmHg]
_
1 to 75
0 to 74
50
– –
FiCO2
[mmHg]
_
0 to 10
– –
5
– –
MV
[L/min]
_
0.1 to 39
0 to 38.9
12
3.0
FiO2
[Vol.%]
_
19 to 100
18 to 99
– –
20
Fi Hal.
[Vol.%]
_
0.1 to 7
0 to 6.9
1.5
– –
Fi Iso.
[Vol.%]
_
0.1 to 7
0 to 6.9
2.3
– –
Fi Enf.
[Vol.%]
_
0.1 to 7
0 to 6.9
3.4
– –
Fi Des.
[Vol.%]
_
0.1 to 21.9
0 to 21.8
12.0
– –
Fi Sev.
[Vol.%]
_
0.1 to 9.9
0 to 9.8
3.4
– –
PAW
[cmH2O]
_
5 to 98
0 to 35
40
8
Configuring in Standby Mode
Setting Default Values
Operating Instructions Julian SW 3.n
97