9
GETTING STARTED
IMPORTANT NOTICE
Charge Battery Before Use
Always charge the battery fully before using the instrument!
The instrument will fully charge its internal battery in three (3)
hours.
The instrument will always operate on the charger and is
designed to do so regardless of the state of charge of the
battery.
Refer to the User Guide for additional information.
Set Time and Date
From the Home screen, press the Setup Instrument
( ) icon,
then press Set Time and Date.
Set the time zone applicable to your application using the Time
Zone drop down selection list.
NOTE: Time and/or date cannot be changed if monitoring is on.
The Xplorer also lets you synchronize time, so that all of your
devices are using the same time for data recording. Time sync
options include GPS, NTP or Real Time Clock (RTC). Each Time
Sync Source button toggles the respective source enabled or
disabled.
If all three are enabled, time sync is sourced by the instrument in
the following priority order: If GPS is available, then it is used. If
not and NTP is available, then NTP is used. If neither GPS nor
NTP is available, then RTC is used.
NOTE: If the battery is removed and only RTC time is available,
the instrument reverts to the default time/date.
Refer to the User Guide for additional information.