DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Small Parts Measurements and Calculations
- 126 -
Figure 10-2 Small Parts Worksheet
The diagnosis editing column displays the cursor “І”, and you can enter diagnosis information.
To print the report:
Press Print in the Small Parts Worksheet.
Printing reference Section 5.8, Printing.
10.3. Others
Select Others to enter another application measurement.
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Urology Measurements and Calculations
- 127 -
Chapter 11 Urology Measurements and Calculations
11.1. Measurement and Calculation
The urology examination is usually in the B mode.
1. Press Exam and select Urology, and then press Set.
2. Press to enter the B mode.
3. Press Measure to activate the measurement function. The system displays the measurement
menu shown as below.
Figure 11-1 Urology Measurement and Calculation Menu
Label Description Method
RUV
Residual Urine Volume
(mL or L)
RUV (mL) = 0.7x RUV-L (mm) × RUV-W (mm) × RUV-H (mm)
/1000
RUV-L
RUV-W
RUV-H
Residual Urine Length
Residual Urine Width
Residual Urine Height
Distance (mm)
PV
Prostate Volume
(mm
3
, cm
3
,
or dm
3
)
PV (mm
3
) = 0.52 x PV-L (mm) × PV-W (mm) × PV-H (mm) /1000
PV-L
PV-W
PV-H
Prostate Length
Prostate Width
Prostate Height
Distance (mm)
PPSA
Predicted Prostate
Specific Antigen Density
PPSA (ng/mL) = 0.12 × PV
SPSA
Serum of Prostate
Specific Antigen
Key in SPSA (ng)
PSAD
Prostate Specific
Antigen Density
PSAD (ng/mL) = SPSA (ng)/ PV (mL), (0.01ng SPSA100ng)
Table 11-1 Urology Measurement and Calculation Items
To determine the residual urine volume or prostate volume, take three measurements: length,
width, and height. The system calculates the volume.

Table of Contents