M3A Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 64 -
Message
Cause
Alarm
Level
What to do
MEASURE
TIMEOUT
Measuring time has
exceeded 120 s (adult)
or 90 s (neonatal).
Low
Measure again or use other
measuring methods.
INIT
PRESSURE
TOO HIGH
The initial pressure is
too high during
measuring.
Low
Measure again, if failure persists,
stop using measuring function of
NIBP module and notify
biomedical engineer or
manufacturer’s service staff.
PRESSURE
RANGE
EXCEED
The measured pressure
exceeds the limit.
Low
Measure by other method.
Prompt message:
Message
Cause
Press NIBP START
You can start NIBP measuring of continual mode.
Manual measuring...
During manual measuring mode.
Continual measuring...
During continual measuring mode.
Automatic measuring...
During automatic measuring mode.
Measurement over
Measurement over
Calibrating...
During calibrating
Calibration over
Calibration over
Leakage testing...
During leakage test
Leakage test over
Leakage test over
NIBP Resetting...
NIBP module is resetting
NIBP Alarm Off
The alarm of NIBP is turned off.