EasyManua.ls Logo

Edwards STP-iXA3306CV - Page 139

Edwards STP-iXA3306CV
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 127
5
SERIAL C
OMMUNICATION PROTOCO
L
Parameter
Item
Data format
Remarks
11
Pump Overload Warning
Rotational speed setting
16-bits hexadecimal
coded ASCII
Default 0000
hex
(0.0%)
12
Serial communication
time out setting (Unit: sec.)
16-bits hexadecimal
coded ASCII
Default 003C
hex
(60 sec.)
13
[System reservation]
88-bits hexadecimal
coded ASCII
*2
*1 Assign the parameter value the reading data of ReadOptionFunc (?=) or FFFF32003CFFhex.
*2 Assign the parameter value the reading data of ReadOptionFunc (?=) or Fhex.
Refer to Table 41, "Parameter setting value" for each parameter setting value.
The value is not reflected when a parameter is out of a setting range.
Example:
1. Input operation port: 01
hex
= I/O REMOTE
2. TMS Option: FF
hex
= DISABLE
4. Second Damage Limit Option: 00
hex
= ENABLE
5. First Damage Limit Warning: 00
hex
= ENABLE
6. Pump Runtime Over Warning: FF
hex
= DISABLE
7. Pump Runtime Over Warning hours: 000003E8
hex
= 1,000 (×100 hours)
8. Imbalance Warning: 00
hex
= ENABLE
9. Pump Overload Warning: FF
hex
= DISABLE
10. Pump Overload Warning Motor current: 03E8
hex
= 1,000 (×0.1%)
11. Pump Overload Warning Rotational speed: 0000
hex
= 0 (×0.1%)
12. Serial communication time out: 003C
hex
= 60 sec.
Parameter
1
2
3
4
5
ASCII
0
1
F
F
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
0
0
0
HEX
30
31
46
46
30
30
30
30
Parameter
6
7
8
ASCII
F
F
0
0
0
0
0
3
E
8
0
0
HEX
46
46
30
30
30
30
30
33
45
38
30
30
Parameter
9
10
11
12
13
ASCII
F
F
0
3
E
8
0
0
0
0
0
0
3
C
*2
*2
*2
*2
*2
*2
HEX
46
46
30
33
45
38
30
30
30
30
30
30
33
43

Table of Contents

Related product manuals