EasyManua.ls Logo

Edwards STP-iXA3306CV - Technical Data; STP Pump Specifications Overview

Edwards STP-iXA3306CV
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 13
TECHNICAL DATA
2
2 TECHNICAL DATA
2.1 STP pump specifications
The values shown below are typical. They are not guaranteed.
Item
STP-iXA3306C(V) / STP-iXA3306B(V)
Flange size
Inlet port flange
VG250 / ISO250F
ICF305 (DN250CF)
VG300 / VG350
ISO320F
ICF356 (DN320CF)
Outlet port flange
KF40
Pumping speed
N
2
L/s
2650
3200
Ar
L/s
2300
2800
H
2
L/s
2050
2300
Compression ratio
N
2,
Ar
>10
8
H
2
>2.0×10
3
Ultimate pressure
Pa (Torr)
10
-7
(10
-9
) [after baking]
Allowable backing pressure
Pa (Torr)
266 (2): Water cooling/TMS unit used
Flow rate of purge gas <N
2
>
Pa·m
3
/s(SCCM)
8.4×10
-2
±1.7×10
-2
(50±10)
Rated speed
rpm
27,700
Backup rotational speed
*1
rpm
Approximately 5,000
Starting time
min
10
Stopping time
min
16
Noise
dB
< 55 (at 27,700 rpm)
Temperature Management System (TMS)
Available
Baking temperature
°C
<120
Lubricating oil
Not necessary
Installation position
Free
Cooling method
Water cooling
Recommended backing-pump
L/min
>1,300
Mass
*2
kg
80
83
Ambient temperature range
°C
0 to 40
Storage temperature range
°C
-25 to 55
*1
A backup rotational speed is the lowest rotational speed to which the magnetic bearing can be backed up at a power failure.
*2
Mass is a value of state that the only standard accessory was installed (except the optional accessory).

Table of Contents

Related product manuals