Device filling/emptying
BA_Elmasteam_8_Vers.07.2016 © Elma Schmidbauer GmbH
6.3.2 Initial Filling with the Canister
When there is no pressure pipe connection with potable water
available, then the pump can also be supplied via an external
canister. To do this, only utilise the original accessories from the
manufacturer.
The commissioning was executed. (Chapter 4 Commissioning)
The device is connected to the canister. (Chapter 4.5 Installation
Water connection with canister)
The device is switched on at the mains supply switch!
The device is ready for use!
Fill the pressure vessel in advance manually with about 2 litres of
water (Chapter 6.2 Manual filling).
Fill the canister with potable water or DI water.
1. The automatic filling or refilling starts approximately 30
seconds after switching on the device.
2.
During the operation, the filling will be regulated via the filling
level monitoring.
3.
The operating display for the filling level flashes when the filling
level of the canister is down to the minimum.
4. Fill the canister with the funnel when the operating display
illuminates. The canister can be refilled during operation.
CAREFUL
Hot vapours/hot liquids
Risk of scalding from open pressure vessel screw cap
• Hot water splashes and evaporates from the filling
opening
• Switch off the device immediately at the mains
switch or unplug the mains supply plug from the
mains supply
• Let the appliance cool down
• Only close the pressure tank screw cap once the