Handling in Operation
72
BA_Elmasteam_8_Vers.07.2016 © Elma Schmidbauer GmbH
CAREFUL
Risk of injury!
Injuries can be caused by chemical burns on the skin!
Never fill chemicals into the device!
Only utilise pure citric acid as 3% in 1.5 litres of
potable water or DI water for descaling processes!
Figure 7.4.5.1: Rotary knob programme selection, service
Utilise the programme service when you want to descale the
device. You require the flushing set and the maintenance hose
(provided accessories) for descaling.
Figure 7.4.5.2: Gauge display descaling 1-8