Discharging the oscillator
10
The clamp is attached to the mounting bracket of
the oscillator
In order to guarantee a safe GND connection
between the crocodile clip and the device ground,
a multimeter is used to ensures that the transition
resistance of the set to the device ground is low
(R<1 Ohm).
To do this, one of the measuring tip on the
multimeter is held on the protective earth of the
device. The other measuring tip is connected to
the tip of the discharge set.
Attention!
The oscillator cover must be removed carefully to
expose the transformer connectors