Faults
51
Fig.
38
: Microcontroller board TP9
NOTICE
When software starts the generator, the voltage on the multimeter should change from 12VDC
to 0.5VDC. The generator is again OFF when this voltage goes again to 12VDC.
The drivers of the microcontroller board are out of order, when the combustion starts without
pressing F5 or clicking START and when without "Analyzing" on the screen, there is only 0,5V
DC or lower on TP9 ( instead of 12V DC ). Replace the electronic drivers
Fig.
39
: Microcontroller board TP13