Rosemount Model 8800C Vortex Flowmeter
4-18
Integral Support Mount (for 6- to 8-in. wafer meters)
6. Remove access cover. See Figure 4-11.
Figure 4-11. Integral Support Tube Assembly
7. Proceed to step 8.
8. Loosen and remove the sensor nut from the sensor cavity with a
1
1
/8-inch open end wrench. (Use a
3
/4-inch open end wrench for 3-
and 4-inch [80 and 100 mm] SST wafers.)
9. Lift the sensor from the sensor cavity. Be very careful to lift the
sensor straight up. Do not rock, twist, or tilt the sensor during
removal; this will damage the engagement diaphragm.
Support Tube
Wafer Meter
Body
Sensor Nut
Sensor
Sensor Cable
Nut
Access Cover
8800-0463A02B