Data Points for dry calibration (continued)Table 5-1:
Daniel MeterLink
Display Name Data points, options and guidelines
Reverse A1 Data points affected:
RevA
1 Options:
Enter a value (dimensionless) within the range [0.95, 1.05].
Guidelines:
This value should only be modified under direction of Daniel
personnel.
Reverse A2 Data points affected:
RevA
2 Options:
Enter a value (s/m or s/ft) within the range [-0.1, 0.1].
Guidelines:
This value should only be modified under direction of Daniel
personnel.
Reverse A3 Data points affected:
RevA
3 Options:
Enter a value s
2
/m
2
or s
2
/ft
2
) within the range [-0.1, 0.1 s
2
/
m
2
].
Guidelines:
This value should only be modified under direction of Daniel
personnel.
Wet calibration gas flow velocity
The Daniel 3410 Series Gas Ultrasonic Flow meter offers three selections for wet
calibration: 12-point piece-wise linearization, a third-order polynomial, or none. The wet
calibration method is selected via the CalMethod data point with "None" being the default
value. The wet calibration gas flow velocity is calculated from the dry calibration gas flow
velocity as shown in Equation 5-2.
Wet-calibration gas flow velocity Equation 5-2:

= 

where
Equation 5-2
V
WetCal
=
wet-calibration gas flow velocity (m/s) (AvgFlow)
V
DryCal
=
dry-calibration gas flow velocity (m/s) (DrycalVel)
Configurations
Operations manual 39