A.5 Environmental specifications
A.5.1 Temperature limits
Ambient temperature
-40 to 70 °C (-40 to 158 °F). Minimum start-up temperature is -50 °C (-58 °F).
With LCD display: -25 to 70 °C (-13 to 158 °F)
Storage temperature
-50 to 85 °C (-58 to 185 °F)
With LCD display: -40 to 85 °C (-40 to 185 °F)
A.5.2
Humidity
0 - 100% relative humidity
A.5.3
Ingress protection
IP 66 and IP 67 (NEMA
®
4X)
A.5.4
Metrology sealing possibility
Yes
A.5.5
Write-protect switch
Yes (hardware and software write-protection)
A.5.6
Transient/built-in lightning protection
In accordance with IEC 61000-4-5, level 4 kV line to ground. Compliant with IEEE 587
category B transient protection and IEEE 472 surge protection.