EasyManua.ls Logo

Emerson Rosemount 2410 - International

Emerson Rosemount 2410
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B.6 International
B.6.1 E7 IECEx Flame-proof
Certificate
IECEx FMG 10.0005X
Standards
IEC 60079-0:2011
IEC 60079-1:2014
IEC 60079-7:2015
IEC 60079-11:2011
Markings
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO and
Ex db eb [ia IIC] IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO and
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity and
Ex db eb [ia IIC] IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC Entity and
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
IP66; IP67
Special Conditions for Safe Use (X):
1. The flamepaths of the equipment are not intended to be repaired. Consult the
manufacturer if repair of the flamepath joints is necessary.
Ex marking
Comm. Bus Ui
V
Ii
mA
Pi
W
Ci
µF
Li
mH
Uo
V
Io
mA
Po
W
Co
µF
Lo
mH
Group
Ex db eb [ib] IIB T4
Gb
FISCO - - - - - 15 354 5.32 - - IIB
Ex db eb [ia IIC Ga]
IIB T4 Gb
HART/4-20mA
Active
- - - - - 23.1 95.3 0.55 0.14 3.9 IIC
1.0 15 IIB
3.67 33 IIA
Ex db eb ib IIB T4
Gb
HART/4-20mA
Passive
30 300 1 0 0 - - - - - IIB
Ex db eb [ib] IIB T4
Gb
Fieldbus - - - - - 15 200 3 1.99 143 IIB
Reference Manual Product certifications
00809-0100-2410 December 2019
Rosemount 2410 Tank Hub 141

Table of Contents

Other manuals for Emerson Rosemount 2410

Related product manuals