B.6 International
B.6.1 E7 IECEx Flame-proof
Certificate
IECEx FMG 10.0005X
Standards
IEC 60079-0:2011
IEC 60079-1:2014
IEC 60079-7:2015
IEC 60079-11:2011
Markings
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO and
Ex db eb [ia IIC] IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC; FISCO and
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity and
Ex db eb [ia IIC] IIB T4 Ta = -50 ºC to 70 ºC Entity
or
Ex db eb [ib] IIB T4 Ta = -50 ºC to 70 ºC Entity and
Ex db eb ib IIB T4 Ta = -50 ºC to 70 ºC Entity
IP66; IP67
Special Conditions for Safe Use (X):
1. The flamepaths of the equipment are not intended to be repaired. Consult the
manufacturer if repair of the flamepath joints is necessary.
Ex marking
Comm. Bus Ui
V
Ii
mA
Pi
W
Ci
µF
Li
mH
Uo
V
Io
mA
Po
W
Co
µF
Lo
mH
Group
Ex db eb [ib] IIB T4
Gb
FISCO - - - - - 15 354 5.32 - - IIB
Ex db eb [ia IIC Ga]
IIB T4 Gb
HART/4-20mA
Active
- - - - - 23.1 95.3 0.55 0.14 3.9 IIC
1.0 15 IIB
3.67 33 IIA
Ex db eb ib IIB T4
Gb
HART/4-20mA
Passive
30 300 1 0 0 - - - - - IIB
Ex db eb [ib] IIB T4
Gb
Fieldbus - - - - - 15 200 3 1.99 143 IIB
Reference Manual Product certifications
00809-0100-2410 December 2019
Rosemount 2410 Tank Hub 141