B.7 Brazil
B.7.1 E2 INMETRO Flame-proof
Certificate
UL-BR 17.1017X
Standards
ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-1:2016, ABNT NBR
IEC 60079-7:2008, ABNT NBR IEC 60079-11:2013
Markings
Ex db e [ib] IIB T4 Gb
Ex db e [ia IIC] IIB T4 Gb
Ex db e ib IIB T4 Gb
Tamb= -50 °C a +70 °C
IP66/IP67
Special Conditions for Safe Use (X):
1. See certificate for special conditions.
Ex marking Comm. Bus Ui
V
Ii
mA
Pi
W
Ci
µF
Li
mH
Uo
V
Io
mA
Po
W
Co
µF
Lo
mH
Group
Ex db e [ib] IIB T4
Gb
FISCO - - - - - 15 354 5.32 - - IIB
Ex db e [ia IIC Ga]
IIB T4 Gb
HART/4-20mA
Active
- - - - - 23.1 95.3 0.55 0.14 3.9 IIC
1.0 15 IIB
3.67 33 IIA
Ex db e ib IIB T4 Gb HART/4-20mA
Passive
30 300 1 0 0 - - - - - IIB
Ex db e [ib] IIB T4
Gb
Fieldbus - - - - - 15 200 3 1.99 143 IIB