B.4 North America
B.4.1 E5 USA Explosion-proof
Certificate
FM16US0123X
Standards
FM Class 3600:2018,
FM Class 3610:2018,
FM Class 3615:2018,
FM Class 3810:2005,
NEMA 250-2003,
ANSI/IEC 60529:2004,
ANSI/UL 60079-0:2013,
ANSI/UL 60079-7:2017,
ANSI/UL 60079-11:2014,
ANSI/UL 61010-1:2004
Markings
FISCO
For b = Tank Bus (Fieldbus - Power and Communication): F and when d =
Secondary Communication Bus (Non-IS): R, E, 5, K, L, V, H, G, A, U, T, B,
6, 7, 0, or F: FISCO POWER SUPPLY XP CL 1, DIV 1 GPS C, D & Associated
Apparatus providing IS circuit to CL I, DIV 1, GPS C & D; DIP CL II/III, DIV.
1, GP E, F & G; CL I, ZONE 1 AEx db eb [ib] IIB Amb. Temp. Limits -50°C to
+70°C Temp. Class T4 SEE CONTROL DRAWING D9240040-901 ENCL.
TYPE 4X, IP66, IP67.
Markings
FISCO HART
active
When b = Tank Bus (Fieldbus - Power and Communication): F and when
d = Secondary Communication Bus (HART
®
/4-20mA Active IS Input/
Output): W, C or 8: FISCO POWER SUPPLY XP CL 1, DIV 1 GPS C, D &
Associated Apparatus providing IS circuit to CL I, DIV 1, GPS C & D; DIP
CL II/III, DIV. 1, GP E, F & G; CL I, ZONE 1 AEx db eb [ib] IIB ENTITY IS I/O
ACTIVE: XP CL 1, DIV 1 GPS C, D & Associated Apparatus providing IS
circuit to CL 1, DIV 1, GPS C & D ACTIVE: CL I, ZONE 0 AEx db eb [ ia IIC ]
IIB Amb. Temp. Limits -50°C to +70°C Temp. Class T4 SEE CONTROL
DRAWING D9240040-901 Type 4X; IP66/67.
Markings
FISCO HART
passive
When b = Tank Bus (Fieldbus - Power and Communication): F and when
d = Secondary Communication Bus (HART
®
/4-20mA Passive IS Input/
Output): D or 9. FISCO POWER SUPPLY XP CL 1, DIV 1 GPS C, D &
Associated Apparatus providing IS circuit to CL I, DIV 1, GPS C & D; DIP
CL II/III, DIV. 1, GP E, F & G; CL I, ZONE 1 AEx db eb [ib] IIB ENTITY IS I/O
PASSIVE: CL I, ZONE 1 AEx db eb ib IIB Amb. Temp. Limits -50°C to +70°C
Temp. Class T4 SEE CONTROL DRAWING D9240040-901 Type 4X;
IP66/67.
Markings
Entity
When b = Tank Bus (Fieldbus - Power and Communication): E and when
d = Secondary Communication Bus (Non-IS): R, E, 5, K, L, V, H, G, A, U, T,
B, 6, 7, 0, or F: ENTITY IS POWER SUPPLY XP CL I, DIV 1 GPS C, D &
Associated Apparatus providing IS circuit to CL I, DIV 1, GPS C & D; DIP
CL II/III, DIV. 1, GP E, F & G; CL I, ZONE 1 AEx db eb [ib] IIB ENTITY Uo:
15.0 V, lo: 200 mA, Po: 3.0 W Co: 1.9 μF, Lo: 143 μH Amb. Temp. Limits
Product certifications Reference Manual