Emerson Process Management GmbH & Co. OHG 1-15
1
Techn. Description
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
1.8 Measurements Specications
Sample gas components and measuring ranges (standard congurations)
In total, more than 60 gases are detectable, so the following table gives an overview only.
Consult with Emerson for gases / congurations not listed.
Not all data is applicable to all analyzer variations. The sample gas(es) and measuring ranges
for your specic analyzer are given by the order acknowledgement and on the analyzer's name
plate label.
Tab. 1-6: Gas Components and Measuring Ranges, examples
Special Specs
or Conditions
Standard Specs
(see Tab. 1-7 1-9)
Enhanced Specs
(see Tab. 1-7 & 1-9)
Gas component Principle
Lowest
Range
Lowest
Range
Lowest
Range
Highest
Range
Acetone
1
CH
3
COCH
3
UV 0–400 ppm 0–800 ppm 0–3 %
Acetone
1
CH
3
COCH
3
IR 0–500 ppm 0–1000 ppm 0–3 %
Acetylene C
2
H
2
IR 0–3 % 0–6 % 0–100 %
Ammonia NH
3
IR 0–100 ppm 0–200 ppm 0–100 %
Argon Ar TCD 0–50 % 0–100 % 0–100 %
Carbon dioxide CO
2
IR 0–5 ppm
5
0–50 ppm 0–100 ppm 0–100 %
Carbon monoxide CO IR 0–10 ppm
5
0–50 ppm 0–100 ppm 0–100 %
Chlorine Cl
2
UV 0–300 ppm 0–600 ppm 0–100 %
Ethane C
2
H
6
IR 0–1000 ppm 0–2000 ppm 0–100 %
Ethanol
1
C
2
H
5
OH IR 0–1000 ppm 0–2000 ppm 0–10 %
Ethylene C
2
H
4
IR 0–400 ppm 0–800 ppm 0–100 %
Helium He TCD 0–10 % 0–20 % 0–100 %
Hexane
1
C
6
H
14
IR 0–100 ppm 0–200 ppm 0–10 %
Hydrogen
4
H
2
TCD 0–1 %
0–2 %
0–100 %
Hydrogen Sulde H
2
S UV 0–2 % 0–4 % 0–10 %
Hydrogen Sulde H
2
S IR 0–10 % 0–20 % 0–100 %
Methane CH
4
IR 0–100 ppm 0–200 ppm 0–100 %
Methanol
1
CH
3
OH IR 0–1000 ppm 0–2000 ppm 0–10 %
n–Butane C
4
H
10
IR 0–800 ppm 0–1600 ppm 0–100 %
Nitrogen dioxide
1
NO
2
UV 0–25 ppm
3
0–50 ppm 0–100 ppm 0–10 %
Nitrogen monoxide NO IR 0–100 ppm 0–200 ppm 0–100 %
Nitrous oxide N
2
O IR 0–100 ppm 0–200 ppm 0–100 %
Oxygen O
2
electrochem. 0–5 % 0–25 %
2 6
Oxygen O
2
paramagn. 0–1 % 0–2 % 0–100 %
Oxygen, Trace O
2
electrochem. 0–10 ppm 0–10 000 ppm
6
Propane C
3
H
8
IR 0–1000 ppm 0–2000 ppm 0–100 %
Propylene C
3
H
6
IR 0–400 ppm 0–800 ppm 0–100 %
Sulfur dioxide SO
2
UV 0–25 ppm
3
0–50 ppm 0–130 ppm 0–1 %
Sulfur dioxide SO
2
IR 0–1 % 0–2 % 0–100 %
Sulfur hexauoride SF
6
IR 0–5 ppm
3
0–20 ppm 0–50 ppm 0–2 %
Toluene
1
C
7
H
8
UV 0–300 ppm 0–600 ppm 0–5 %
Vinyl chloride C
2
H
3
Cl IR 0–1000 ppm 0–2000 ppm 0–2 %
Water vapor
1
H
2
O IR 0–1000 ppm 0–2000 ppm 0–8 %
Water vapor, Trace
1
H
2
O capacitive 0–100 ppm 0–3000 ppm
6
1
Dew point below ambient
temperature
2
Higher concentrations
decrease sensor lifetime
3
Daily zero calibration re- qui-
red for ranges below lowest
standard specs range
4
Special "renery" applica-
tion with 0–1% H
2
in N
2
available
5
see Tab. 1-10
6
standard specs only
1.8 Measurements Specications

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals