Modications made on the electrical appara-
tus affecting the integrity of the type of protec-
tion or the temperature of the apparatus shall
be permitted only if the modied apparatus is
resubmitted to a testing station.
4.1 Verications and Tests
4.1 Verications and Tests
4.1.1 Routine Tests
The following tests shall be performed on a
regular basis:
EXPLOSION HAZARD BY DAMAGED ANALYZER
Do not continue to operate a damaged analyzer!
Call for service!
Violation may cause an explosion and/or personal injury or death!
4.1.1.1 Visual Inspection
The instrument shall be visually checked for
damages to the enclosure, ange, threads
and external analyzer components.
Especially take care that there are no scrat-
ches on the anges surfaces!
Contents
4.1 Verications and Tests ........................................4-3
4.2 Replacement of Parts .........................................4-6
4.3 Vapor Recovery Application - Special Information ...................4-8
4.4 Perform a calibration ........................................4-15
In the case of repairs affecting the type of
protection, the parts which have been re-
paired should be subjected to new routine
verications and tests. These tests need not
necessarily be made by the manufacturer.
4.1.1.2 Detailed Inspection
Check cable inlets (glands or conduits): If
the cable moves, e. g. thighten the cap nut.
If tightening is ineffective, replace the cable
gland or conduit.
Also consider the original manufacturer‘s
instructions for maintenance!

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals