Emerson Process Management GmbH & Co. OHG 1-5
1
Techn. Description
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
1.5 Compliances
1.5.1 Special Conditions for Safe Use
Only screws M16x45 ISO 4762 A2-70 as
specied in the maintenance section of
this manual shall be used (spare part #
42716945).
The ame joints correspond to the drawing
No. 4.271-7112/1 and do not comply with
the dimensions mentioned into the Tab. 1
and Tab. 2 of EN 60079-1 ed.2.
The gas path for the sample gas shall be
equipped with additional appropriate ame
arrestors in case of gas pressure above
1100 hPa to 1500 hPa.
Appropriate certied cable glands shall be
used in accordance with IEC/EN 60079-14.
Vapor Recovery application:
Pressure of gases not to exceed 1100 hPa.
Concentration of gases must be below
25 % LEL.
Depending on the particular application all
approbriate safety instructions mentioned
in this instruction manual on hand must
be considered!
Take special care of formation of amma-
ble gas at the outlet of breathing and/or
purging devices, if the sample gas con-
centration is above 25% LEL! If need be,
such outlets have to end in a safe area!

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals