Emerson Process Management GmbH & Co. OHG 4-5
4
Maintenance
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
Fig. 4-1: Leak Testing with U-tube Manometer (Flame arrestor)
Analyzer
overpressure approx.
0.725 psi/50 mbar
stop
valve
Water
ame
arrestors
4.1.1.3.2 Pressure Drop Test
To measure the pressure drop at a ame
arrestor
disconnect internal and external gas lines
connect a ow meter in series to the ame
arrestor
connect an air or nitrogen supply to the
ow meter
connect a manometer in a way to measu-
re the pressure at the ame arrestor inlet
against the outlet
• adjust the ow to 1 l/min and compare the
resulting pressure drop to the following
reference value: (5.1 ± 1) hPa.
Replace the ame arrestor if the measured
pressure value differs more than 10 % from
the reference value given above.
Fig. 4-2: Pressure Drop Test
Q = 1 l/min
ame
arrestor
ow meter
manometer
4.1 Verications and Tests

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals