EasyManuals Logo

Emerson Rosemount Xstream Series Instruction Manual Addendum

Emerson Rosemount Xstream Series
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Emerson Process Management GmbH & Co. OHG 1-9
1
Techn. Description
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
1.7 Technical Data
Standard interfaces
up to 5 analog outputs
(standard: 1 analog output
per channel)
electrical
specication
4 (0)–20 mA (R
B
500 )
optically isolated from each other and from analy-
zer electronics
function
user-congurable activation and deactivation of
concentration levels
support for NAMUR NE 43 operation modes, con-
gurable via keypad and Modbus
4 relay outputs
electrical
specication
dry contacts
max. load. 30 V; 1 A; 30 W resistive
function
Each output can be congured to provide any of
the functions listed in chapter 6 of the X-STREAM
XE instruction manual
.
These functions include,
but are not limited to
NAMUR NE 107 status signals 'Failure', 'Main-
tenance request', 'Out of specication', 'Function
check',
concentration alarms,
control signals for external valves or pumps,
and many more
2 Modbus interfaces Ethernet (RJ45 sockets), 10/100 MBit
2 USB ports
specication USB 1.0
function
1 USB connector type A,
for connecting external storage devices
1 USB connector type mini AB,
for connecting external computers
Tab. 1-2: Standard Interfaces Data

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals