1.8 Measurements Specications
Special Performance Specications for Gas Purity Measurements (ULCO & ULCO
2
)
Tab. 1-10: Special Performance Specications for Gas Purity Measurements
1
Related to full scale
2
Constant pressure and temperature
3
Within 24 h; daily zero calibration requested
4
Within 24 h; daily span calibration recommended
5
Related to measuring value
6
Temperature variation: 10 K per hour
7
From gas analyzer inlet at gas ow of 1.0 l/min
8
Barometric pressure sensor is required
9
Whichever value is higher
10
Limited to atmospheric if internal sample pump
11
Special conditions apply to pressures above
1100 hPa
(1.5 psig) up to 1500 hPa (7 psig)
0–10…< 50 ppm CO
0–5…< 50 ppm CO
2
Detection limit (4 σ)
1 2
<
2 %
Linearity
1 2
< 1 %
Zero-point drift
1 2 3
< 2 % resp. < 0.2 ppm
9
Span (sensitivity) drift
1 2 4
< 2 % resp. < 0.2 ppm
9
Repeatability
1 2
< 2 % resp. < 0.2 ppm
9
Response time (t
90
)
7
< 10 s
Permissible gas ow 0.2–1.5 l/min.
Inuence of gas ow
1 2
< 2%
Maximum gas pressure
10 11
≤ 1500 hPa abs.
( 7 psig)
Inuence of pressure
5
At constant temperature 0.1 % per hPa
– With pressure compensation
8
0.01 % per hPa
Permissible ambient temperature +15 to +35 °C (59 to 95 °F) +5 to +40 °C (41 to 104 °F)
Inuence of temperature
6
(at constant pressure)
– On zero point < 2 % per 10 K resp. < 0.2 ppm per 10 K
9
– On span (sensitivity) < 2 % per 10 K resp. < 0.2 ppm per 10 K
9
Thermostat control none 60 °C (140 °F)
Note! 1 psi = 68.95 hPa

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals