P100 USERS MANUAL JUN2014v03
54
A.2 Materials
EMTEK P100 Air Sampler Calibrated for operation at 28.3 LPM (1 CFM), or 100 LPM with
applicable inlet cover.
100 LPM Inlet Cover Blue Anodized 28.3 LPM (1CFM) Inlet Cover Clear Anodized
NOTE: If exhaust tubing will be used with the unit, it is ideal that the unit is calibrated with the
exhaust tubing in place.
Standard 90 mm TSA Agar Test Plates
30-35°C and/or 20-25°C Incubators (or other appropriate temperature).
Gloves (Sterile or Clean)
Low Particulate Shedding Wipes (e.g., Wipe All, Gamma Wipes or equivalent/better)
Disinfectants (e.g., Quaternary Amonium Compounds, 70% 0.2µm Filtered or sterilized alcohol)
Lab Marker and/or Pre-Printed Labels
Adhesive Tape
A.3 Maintenance Inspection
To assure appropriate operation of the P100 sampling assembly, prior to each days use, sampling personnel should
inspect the unit for any obvious physical defect. This inspection shall include but not be limited to a visual
inspection of the sample inlet cover inlet holes to assure that they are free of occlusions to assure proper sample
flow.
If any maintenance need be performed, contact appropriate service or repair personnel.