Appendix B: Suggested P100 Sanitization Procedure
B.1 Materials
• Gloves (Sterile or Clean)
• Low Particulate Shedding Wipes (e.g., Wipe All, Gamma Wipes)
• Primary Disinfectants (e.g., Quaternary Ammonium Compounds)
Important Note: Phenolic disinfectants should not be used on the P100 Enclosure, or touchscreen overlay, as it
will cause the components to become very brittle in a short period of use.
• Secondary Disinfectants (e.g., 70% 0.2µm filtered or sterilized alcohol)
• PLUGS: Blower inlet plug, Ethernet port plug, USB port plug
B.2 Sanitization Procedure
English
DANGER: TO MINIMIZE THE CHANCE OF ELECTRIC SHOCK disconnect the P100
AC/DC power supply cord, and power off the unit during sanitization
DANGER: DO NOT SUBMERSE the P100 Air Sampler in any liquids!
Français
DANGER: AFIN DE RÉDUIRE LES RISQUES DE CHOC ÉLECTRIQUE débrancher le P100
Cordon d'alimentation AC / DC, et l'appareil hors tension pendant la désinfection
DANGER: NE PAS immerger le contrôleur de P100 Air Samplre dans tous les liquides!