LINE 3: ADMIN OPTIONS: Allows access to ADMIN CONTROL screen. Requires entry of ADMIN
ID and Password to access ADMIN CONTROL screen.
Administrative Control Screen Options
Line 1 PROGRAMS: Controls whether or not the USER can create or delete Saved Programs
(OFF=User, ON=Admin Only)
Line 2 CLEAR DATA: Controls whether or not the user can clear stored data
(OFF=User, ON=Admin Only).
DATE STORAGE: The P100 will store up to 500 samples in its internal memory. A
warning will occur at power up, which requires acknowledgement, when memory hits 450
samples. The CPU will delete oldest sample runs to add new ones once 500 samples has been
stored in the memory, or will prevent new runs from being taken once 500 runs have been
stored if the CLEAR DATA function is ON, preventing data runs from being deleted.
Line 3 SITE DESC: Controls whether or not the user can add or delete site descriptions
(OFF=User, ON=Admin Only)
Line 4 CALIB DUE ALM: Controls whether or not the user can turn the calibration due alarm
ON or OFF (OFF=User, ON=Admin Only). If set to ON by ADMIN will set notification
ON on the CAL screen as well.
Line 5 FLOW ALM: Controls whether or not the user can turn the calibration due alarm ON or
OFF (OFF=User, ON=Admin Only)
NOTE: Once the options are set as desired, the P100 must be power cycled (Turned OFF then ON) to log
out the ADMIN and lock these functions as set, otherwise the can still be modified (turned ON or
OFF) by the user.