P100 USERS MANUAL JUN2014v03
50
7.5 Replacing the HEPA Filter and/or Battery Pack
IMPORTANT NOTE: To replace the HEPA filter, or battery pack, contact EMTEK, or your local area service
representative. Only certified service technicians should perform a filter, or battery
replacement.
NOTE IMPORTANTE: Pour remplacer le filtre HEPA, ou la batterie, contactez EMTEK, ou votre représentant du
service local. Seuls les techniciens de maintenance certifié doit effectuer un remplacement
du filtre, ou la batterie.
HEPA FILTER: The P100 is equipped with a HEPA filter exhaust. It is recommended to replace the HEPA
filter every 12 months as part of periodic maintenance.
BATTERY PACK: The Lithium Ion battery pack may need replacement within 2-3 years.
NOTE: When replacing a battery pack, follow the steps below to reset the battery life indicator:
1) Replace the battery pack (performed by trained technician!).
2) Power on the P100
3) Fully discharge the P100. This is best achieved by entering a sample volume of “0”, and a flow rate of
100 LPM. This will run the battery down, without multiple interactions by the user.
4) Once the battery life is down to “0”, or the LOW BATTERY indicator has popped up, plug the
charger in to the unit. The unit will charge faster if not powered on.
5) After 2.5 hours, check the battery life indicator to see if the indicator shows 100%. If so, the P100 is
ready for use. It may take two full discharge charge cycles to get maximum battery life out of the
P100.