P100 USERS MANUAL JUN2014v03
59
1.7 Ending the Testing Period:
1.7.1 When the required volume is attained the blower motor automatically shuts off. If desired
the test period can be terminated by pressing the Stop key on the P100, or on the IR remote.
1.7.2 If desired, the test period can be Pause/Resume by pressing the Pause/Resume key on the
P100 or IR remote control at any time during the testing period.
1.8 Aseptically remove the test plate as follows:
NOTE: Gloved hands should be rinsed with secondary disinfectant and allowed to dry immediately
prior to performing these steps.
1.8.1 Remove the inlet cover by turning it counter clockwise to unlock it. Hold the inlet cover
above the test plate with one hand (Fig B), and with the other hand replace the lid of the test
plate (Fig C). Be careful not to touch the inside of the lid or the agar surface. Do not move
your hand over the exposed surface of the test plate. Allow the lid to lead your hand over
the test plate as you replace it.
Fig A Fig B Fig C
1.8.2 Remove the test plate, with the lid in place (Fig A), from the base plate holder and replace the
P100 inlet cover (Fig B).
Fig A Fig B
1.8.3 Secure the test plate lid to the test plate with tape.