Deltabar S
Endress+Hauser 5
Measuring principle
Ceramic process isolating diaphragms used for PMD70 and FMD76
The ceramic measuring cell is based on the principle of a plate capacitor with an electrode on (1) and a movable
electrode on the interior of the diaphragm (3). Standard silicone oil or mineral oil filling oils for this measuring
cell.
A differential pressure (p
1
p
2
) causes a corresponding deflection of both diaphragms. Both capacitance values
are converted and are fed to the microprocessor of the transmitter as a digital signal.
Advantages:
Self-monitoring for process isolating diaphragm break or oil loss (constant comparison of the measured
temperature with a temperature calculated from the capacitance values)
Extremely high resistance to aggressive media
Suitable for vacuums up to 1 mbar
abs
Metal-free versions available
Second process barrier (Secondary Containment) for enhanced integrity
Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
The process isolating diaphragms (3/9) are deflected on both sides by the acting pressures. A filling oil (4/8)
transfers the pressure to a resistance circuit bridge (semi-conductor technology). The differential-pressure-
dependent change of the bridge output voltage is measured and further processed.
Advantages:
Standard operating pressures: 160 bar and 420 bar
High long-term stability
Very high single-sided overload resistance
Second process barrier (Secondary Containment) for enhanced integrity
Ceramic process isolating diaphragms used for
PMD70 and FMD76
Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
P01-xMD7xxxx-03-xx-xx-xx-000
Ceramic measuring cell PMD70 and FMD76
1 Meter body
2 Process isolating diaphragm
3 Electrodes
4 Glass frit fixes the process isolating
diaphragm onto the meter body
5 Temperature sensor
P01-xMD7xxxx-03-xx-xx-xx-002
Metal measuring cell 10 mbar and 30 mbar
1 Sensing element
2 Silicon diaphragm
3 Process isolating diaphragm
4 Filling oil
5 Overload diaphragm
P01-xMD7xxxx-03-xx-xx-xx-003
Metal measuring cell as of 100 mbar
6 Sensing element
7 Overload diaphragm
8 Filling oil
9 Process isolating diaphragm
1
p
2
p
1
p
2
p
1
p
2
p