EasyManua.ls Logo

Endress+Hauser Deltabar S PMD75 - Local Operation; Remote Operation

Endress+Hauser Deltabar S PMD75
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Deltabar S
Endress+Hauser 55
Local operation
Remote operation Depending on the position of the write protection switch at the device, all software parameters are accessible.
HART
Remote operation via:
Handheld terminal Field Communicator 375 (see Chapter "Hard- und Software for on-site and remote
operation" ä 56)
FieldCare (see Chapter "Hard- und Software for on-site and remote operation" ä 56 ff) mit
Commubox FXA191 (see Chapter "Hard- und Software for on-site and remote operation" ä 56 ff)
Commubox FXA195 (see Chapter "Hard- und Software for on-site and remote operation" ä 56 ff)
Field Xpert:
Field Xpert is an industrial PDA with integrated 3.5" touchscreen from Endress+Hauser based on Windows
Mobile. It communicates via wireless with the optional VIATOR Bluetooth modem connected to a HART
device point-to-point or wireless via WiFi and Endress+Hauser’s Fieldgate FXA520. Field Xpert also works
as a stand-alone device for asset management applications. For details refer to BA060S/00/en.
PROFIBUS PA
Remote operation via:
FieldCare (see Chapter "Hard- und Software for on-site and remote operation" ä 56 ff)
Profiboard: For the Connection of a Personal Computer to PROFIBUS
Proficard: For the Connection of a Laptop to PROFIBUS
FOUNDATION Fieldbus
Remote operation via:
Handheld terminal Field Communicator 375 (see Chapter "Hard- und Software for on-site and remote
operation" ä 56 ff)
Use an FF-configuration program for e.g. NI-FBUS configurator, to
connect devices with "FOUNDATION Fieldbus signal" into an FF-network
set FF-specific parameter
Operation with NI-FBUS Configurator:
The NI-FBUS Configurator is an easy-to-use graphical environment for creating linkages, loops and a
schedule based on the fieldbus concept.
You can use the NI-FBUS Configurator to configure a fieldbus network as follows:
Set block and device tags
Set device addresses
Create and edit function block control strategies (function block applications)
Configure vendor -defined function and transducer blocks
Create and edit schedules
Read and write to function block control strategies (function block applications)
Invoke Device Description (DD) methods
Function External operation
(operation keys, optio-
nal, not T17 housing)
Internal operation
(electronic insert)
Display (optional)
Position calibration
(zero point correction)
XXX
Setting lower-range value
and upper-range value -
reference pressure present
at the device
X
(HART only)
X
(HART only)
X
Device Reset X X X
Locking and unlocking
measured-value-relevant
parameters
XX
Value acceptance indica-
ted by green LED
XXX
Switching damping on and
off
X
(HART and PA only)
X
Setting bus address (PA) XX
Switching simulation
mode on and off (FOUN-
DATION Fieldbus)
XX

Table of Contents

Other manuals for Endress+Hauser Deltabar S PMD75