Deltabar S
54 Endress+Hauser
Operating keys on the exterior of the device
The operating keys of the housing T14 (aluminium or stainless steel) are located either outside of the housing,
under the protection cap or upon the electronic insert. The operating keys of the housing T17 (ironing stainless
steel) are located inside the housing upon the electronic insert. In addition, devices with an on-site display and
a 4 to 20 mA HART- or PROFIBUS PA electronic insert have operating keys on the on-site display.
P01-PMx7xxxx-19-xx-xx-xx-038
The operating keys located externally on the device work on the Hall sensor principle. As a result, no additional
openings are required in the device. This guarantees:
Complete protection against environmental influences such as moisture and contamination
Simple operation without any tools
No wear.
Operating keys and elements located internally on the electronic insert
0%
Zero
4...20 mA HART
PROFIBUS PA/
FOUNDATION Fieldbus
P01-xxxxxxxx-19-xx-xx-xx-104
Electronic insert HART
1Operating keys
2 Slot for optional display
3 Slot for optional HistoROM
®
/M-DAT
4 DIP-switch for locking/unlocking
measured-value-relevant parameters
5 DIP-switch for damping on/off
6 Green LED to indicate value being accepted
P01-xxxxxxxx-19-xx-xx-xx-105
Elektronikeinsatz PROFIBUS PA
1 Green LED to indicate value being accepted
2 Key for position calibration and device reset
3 DIP-switch for bus address
4 Slot for optional display
5 Slot for optional HistoROM
®
/M-DAT
6 DIP-switch for locking/unlocking
measured-value-relevant parameters
7 DIP-switch for damping on/off
P01-xxxxxxxx-19-xx-xx-xx-106
Elektronikeinsatz FOUNDATION Fieldbus
1 Green LED to indicate value being accepted
2 Key for position calibration and device reset
3 Slot for optional display
4 Slot for optional HistoROM
®
/M-DAT
5 DIP-switch for locking/unlocking
measured-value-relevant parameters
6DIP-switch for simulation mode on/off
HW-Version:
SW-Version:
250002271-–
HART
R
FIELD COMMUNICATIONPROTOCOL
1
2
21
PC
t
on
off
Display
Sensor
on
off
0%
Zero
Address
SW
HW
1 2
3
4
5
6 78
on
off
His
to
RO
M
on
off
➀➁
t
PC
21
21
CKON
3 4 5 6 78
S D A 08
HW-Version:
SW-Version:
250002272-
1
2
HW
21
PC
1
2
➀➁
on
off
Sim.
Sensor
on
off
Simulation
0%
Zero
Display
His
to
RO
M
HW-Version:
SW-Version:
250002273-
TM
FOUNDATION
1
2

Table of Contents

Other manuals for Endress+Hauser Deltabar S PMD75

Related product manuals