Deltabar S
Endress+Hauser 57
HistoROM
®
/M-DAT (optional)
HistoROM
®
/M-DAT is a memory module, which is attached to the electronic insert. The HistoROM
®
/M-DAT
can be retrofitted at any stage (Order number: 52027785).
Your benefits
Quick and safe commissioning of the same measuring points by copying the configuration data of one
transmitter to another transmitter
Reliable process monitoring thanks to cyclical recording of pressure and sensor temperature measured values
Simple dagnosis by recording diverse events such as alarms, configuration changes, counters for measuring
range undershoot and overshoot for pressure and temperature as well as user limit overshoot and undershoot
for pressure and temperature etc.
Analysis and graphic evaluation of the events and process parameters via software (contained in scope of
supply).
HistoROM
®
/M-DAT can be ordered via feature 100 "Additional options 1" or feature 110 "Additional options
2" or as spare parts. ä 70 ff. A CD with Endress+Hauser operating program is also included in the scope of
delivery.
You can copy data from one transmitter to another transmitter when operating a FOUNDATION Fieldbus
device via an FF configuration program. You need the Endress+Hauser FieldCare operating program and the
Commubox FXA291 service interface and the ToF Adapter FXA291 to be able to access the data and events
saved in the HistoROM
®
/M-DAT.
FieldCare
FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can
configure all Endress+Hauser devices as wella s devices from other manufacturers that support the FDT
standard.
FieldCare supports the following functions:
Configuration of transmitter in offline and online operation
Loading and saving device data (upload/download)
HistoROM
®
/M-DAT analysis
Documentation of the measuring point
Connection options:
HART via Commubox FXA191 and the RS232C serial interface of a computer
HART via Commubox FXA195 and the USB port on a computer
PROFIBUS PA via segment coupler and PROFIBUS interface card
FOUNDATION Fieldbus via Commubox FXA193 and the RS232C serial interface of a computer
Service interface with adapter Commubox FXA291 and ToF Adapter FXA291 (USB).
For further information see È www.endress.com

Table of Contents

Other manuals for Endress+Hauser Deltabar S PMD75