EasyManua.ls Logo

Endress+Hauser iTEMP TMT82 - Page 101

Endress+Hauser iTEMP TMT82
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iTEMP TMT82 Operating menu and parameter description
Endress+Hauser 101
13.3.3 "Output" submenu
Measuring mode
Navigation Expert → Output → Measuring mode
Description Enables the inversion of the output signal.
Additional information Standard
The output current increases with increasing temperatures
inverted
The output current decreases with increasing temperatures
Options Standard
inverted
Factory setting Standard
13.3.4 "Communication" submenu
"HART configuration" submenu
Device tag →  88
Navigation Diagnostics → Device information → Device tag
Expert → Communication → HART configuration → Device tag
HART short tag
Navigation Expert → Communication → HART configuration → HART short tag
Description Use this function to define a short tag for the measuring point.
User entry Up to 8 alphanumeric characters (letters, numbers and special characters)
Factory setting SHORTTAG
HART address
Navigation Expert → Communication → HART configuration → HART address
Description Use this function to define the HART address of the device.